P2000 The voltage of the solar installations (also known as photovoltaic or PV installations) increases to reduce cost and improve the efficiency of solar equipment. As a result, it is important to select measuring instruments that support higher voltage measurement to protect the safety of inspection workers.
The P2000 enables high voltage measurement up to CAT III 2000 V by simply connecting to a compatible Hioki clamp meter or digital multimeter (DMM).
PROBES
HIOKI P2000 DC High Voltage Probe
SPECIFICATIONS
- Maximum Input Voltage : DC 2000 V
- Maximum rated line-to-ground voltage
1000 V (Measurement Category IV), Anticipated transient overvoltage 12000 V
2000 V (Measurement Category III), Anticipated transient overvoltage 15000 V - Input Resistance : 20 MΩ ±5.0% (between INPUT H and INPUT L, when OUTPUT terminal is open)
- Overload Protection :
2200 V DC/2200 V AC (applied for 1 minute) (between INPUT H and INPUT L)
600 V DC/600 V AC (applied for 1 minute) (between OUTPUT H and OUTPUT L) - Output Terminal : 4 mm banana terminal
- Operating Temperature Range : -25°C to 65°C (-13°F to 149°F)
- Operating Humidity Range : −25°C to 40°C (−13°F to 104°F): 80% RH or less (non-condensing)
40°C to 65°C (104°F to 149°F): Linearly reduces from 80% RH or less at 40°C (104°F) to 25% RH or less at 65°C (149°F) (non-condensing) - Storage temperature and humidity range : -30°C to 70°C (-22°F to 158°F) 90% RH or less (non-condensing)
- Standards : EN 61010
- Included Accessories : Connection Cable Set L4943 ×1, Strap with buckle ×2, Strap ×1, Carrying Case C0205 ×1, Instruction Manual ×1, Operating Precautions ×1
FEATURES
- Enables high voltage measurement up to CAT Ⅲ 2000 V
- Safely measure solar installations like the mega-solar without interrupting the grid
- Connects to compatible Hioki clamp meters and digital multimeters (DMMs)
ACCURACY FOR COMPATIBLE INSTRUMENTS
- DT-4261 :
Range 600.0 V: Combined accuracy ±0.5% rdg ±0.2 V
Range 2000 V: Combined accuracy ±0.5% rdg ±5 V - CM4141-50 ; CM4371-50 ; CM4373-50 ; CM4375-50 :
Range 600.0 V: Combined accuracy ±1.0% rdg ±0.3 V
Range 2000 V: Combined accuracy ±1.0% rdg ±3 V - DT-4282 ; DT-4281 :
Range 60.000 V: Combined accuracy ±0.8% rdg ±0.002 V (output rate 1/10)
Range 600.00 V: Combined accuracy ±0.8% rdg ±0.02 V (output rate 1/10) - DT-4252 ; DT-4253 :
Range 60.00 V: Combined accuracy ±1.2% rdg ±0.05 V (output rate 1/10)
Range 600.0 V: Combined accuracy ±1.2% rdg ±0.5 V (output rate 1/10) - DT-4255 ; DT-4256 :
Range 60.00 V: Combined accuracy ±1.2% rdg ±0.03 V (output rate 1/10)
Range 600.0 V: Combined accuracy ±1.2% rdg ±0.3 V (output rate 1/10)
Compare